OBJECTIVE: Cone beam computed tomography (CBCT) images are being increasingly used to acquire three-dimensional (3D) models of the skull for additive manufacturing purposes. However, the accuracy of such models remains a challenge, especially in the orbital area. The aim of this study is to assess the impact of four different CBCT imaging positions on the accuracy of the resulting 3D models in the orbital area. METHODS: An anthropomorphic head phantom was manufactured by submerging a dry human skull in silicon to mimic the soft tissue attenuation and scattering properties of the human head. The phantom was scanned on a ProMax 3D MAX CBCT scanner using 90 and 120 kV for four different field of view positions: standard; elevated; backwards tilted; and forward tilted. All CBCT images were subsequently converted into 3D models and geometrically compared with a "gold-standard" optical scan of the dry skull. RESULTS: Mean absolute deviations of the 3D models ranged between 0.15 ± 0.11 mm and 0.56 ± 0.28 mm. The elevated imaging position in combination with 120 kV tube voltage resulted in an improved representation of the orbital walls in the resulting 3D model without compromising the accuracy. CONCLUSIONS: Head positioning during CBCT imaging can influence the accuracy of the resulting 3D model. The accuracy of such models may be improved by positioning the region of interest (e.g. the orbital area) in the focal plane (Figure 2a) of the CBCT X-ray beam.

, , , ,
Planmeca Oy, Helsinki, Finland
doi.org/10.1259/dmfr.20220104
Dentomaxillofacial Radiology
Centrum Wiskunde & Informatica, Amsterdam (CWI), The Netherlands

van Eijnatten, M., Wolff, J., Pauwels, R., Karhu, K., Hietanen, A., Der Sarkissian, H., & Koivisto, J. (2022). Influence of head positioning during cone-beam CT imaging on the accuracy of virtual 3D models. Dentomaxillofacial Radiology, 51(7), 20220104:1–20220104:9. doi:10.1259/dmfr.20220104